NK Optik Otoflash G171-6 REPLACEMENT Flash Bulb Module (Lasts approx. 250-300 working hours) – REF 171.09-LM – 1pc
This is a replacement part for the Otoflash G171-6 flash-curing device.
The flash module has 2 flash-light bulbs arranged below the device @ 100 watts. Lasts approx. 250-300 working hours.
*** SEE INSTRUCTION MANUAL PAGES 9-10 “Exchanging the flash-module (Fig. 3)” for instructions how to change lamp module. ***
DANGER
Before carrying out any service or maintenance work, the device should be disconnected from the power supply.
– The flash-bulbs can only be replaced by replacing the entire flash-module.
Danger due to hot surface
– Allow hot flash-module to cool off by running the device in stand-by mode for approximately 1 minute.
– Turn off the device and disconnect from power supply.
– Open lid.
– Remove the 4 screws at the corners of the flash-module (Fig. 3/B). Extract the flash-module by grasping it at the centre-front and gently wiggling it, and pulling it up and out from the device.
You may find using a fl athead screwdriver as a lever to be helpful during the removal process.
To do so, carefully wedge the screwdriver in the middle between casing and module and apply gentle pressure to lift up the module.
Carefully insert the new flash- module onto the contact carrier and replace the 4 screws.
Remove the protective foil.
NK Optik Otoflash G171-6 REPLACEMENT Flash Bulb Module (Lasts approx. 250-300 working hours) – REF 171.09-LM – 1pc
- Brand: NK Optik
- Product Code: NK-OLM
NK Optik Otoflash G171-6 REPLACEMENT Flash Bulb Module (Lasts approx. 250-300 working hours) – REF 171.09-LM – 1pc
This is a replacement part for the Otoflash G171-6 flash-curing device.
The flash module has 2 flash-light bulbs arranged below the device @ 100 watts. Lasts approx. 250-300 working hours.
*** SEE INSTRUCTION MANUAL PAGES 9-10 “Exchanging the flash-module (Fig. 3)” for instructions how to change lamp module. ***
DANGER
Before carrying out any service or maintenance work, the device should be disconnected from the power supply.
– The flash-bulbs can only be replaced by replacing the entire flash-module.
Danger due to hot surface
– Allow hot flash-module to cool off by running the device in stand-by mode for approximately 1 minute.
– Turn off the device and disconnect from power supply.
– Open lid.
– Remove the 4 screws at the corners of the flash-module (Fig. 3/B). Extract the flash-module by grasping it at the centre-front and gently wiggling it, and pulling it up and out from the device.
You may find using a fl athead screwdriver as a lever to be helpful during the removal process.
To do so, carefully wedge the screwdriver in the middle between casing and module and apply gentle pressure to lift up the module.
Carefully insert the new flash- module onto the contact carrier and replace the 4 screws.
Remove the protective foil.